OPERATION: 1. Pass one of the detonator leg wire or connecting wire through the DETOMET box 2. Connect the optical cable from the DETOMET to the "Start" receiver of the EXPLOMET-FO 3. Fix an optical cable to the detonator's end with a tape 4. Connect the other end to the "stop" plug of the EXPLOMET-FO 5. Turn the power switch "ON" 6. Press the "RESET" button, the green light flashes 7. Follow the instruction manual of your EXPLOMET-FO instrument 8. Fire the detonator ! 9. The red light flashes, indicating that the DETOMET has detected the firing impulse 10. Read on the EXPLOMET-FO display the delay in microseconds between the firing impulse and the base charge explosion. To start a new measure press the "RESET" button.
Change the battery when the red light is steady or when there is no light at all. To change the battery: remove the back panel of the case, take care of the cable, pull smoothly the battery backwards and turn it 90 degrees to release it from the clips. Caution: Do not remove any other part of the case. SPECIFICATIONS: Dimension: solid aluminum case: 110 x 113 x 60 mm Weight: 0,4 kg Fiber optic cable 1/2,2 mm with connectors for connection with EXPLOMET-FO: 1 m or 20 m. Battery 9V, 500mAh, Block AM6 or 6LR61(UCAR-Alkali) Test lamps on the front panel: - green flashing: Standby- red flashing: shows that the DETOMET has detected the firing |